1. Charge your sensor 

Connect the supplied USB cable to the sensor's mini-USB port and a powered USB port or USB mains charger rated at 5 V, 500 mA or more.

  • Allow up to 4 hours for a full charge.
  • A flashing white status light indicates charging while the sensor is on.
  • A solid green light indicates the battery is fully charged.

2. Switch on and take a reading 

  1. Press the power button on the front of the sensor to switch it on.
  2. Check that the status light flashes blue.

3. Connect to EasySense  

Wireless connection — Bluetooth

  1. Open the EasySense app on your compatible computer, tablet or mobile device.
  2. Switch on the sensor and check that its status light flashes blue.
  3. Select the Devices icon in EasySense.
  4. Find your sensor using its unique ID number.
  5. Select Connect next to your sensor.
  6. Check that the Devices icon turns green and the sensor's status light flashes green.


Important: Do not pair the sensor through your device's Bluetooth settings. EasySense discovers and connects to the sensor directly.

Wired connection — USB

  1. Connect the sensor to your computer using the supplied USB cable.
  2. Open EasySense.
  3. Check that the Devices icon turns green and the sensor indicates a connection.
  4. Begin collecting data.

4. Start your investigation 

Once connected:

  • View live current and voltage readings.
  • Collect and analyse data in EasySense.
  • Save and retrieve your results.
  • Use Remote Mode to record data independently after configuring and starting the recording in EasySense.

For full instructions on data logging, refer to the main manual and the EasySense user guide.

5. Look after your sensor  

Safety check

Limit or instruction

Maximum Voltages and Currents

(measurements cannot be made up to these values)

Maximum voltage on voltage channel: ±32 V

Maximum current on current channel: ±5 A

Maximum voltage on current channel: ±13 V

Electronics operating temperature


0°C to 40°C

Water protection

The electronics housing is not waterproof

Cleaning

Wash the stainless steel probe; wipe the electronics housing with a damp cloth only



6. Quick troubleshooting  

Problem

What to try

No status light

Briefly press the power button.


Sensor not visible in EasySense


Check that it is switched on and the light flashes blue.

Sensor will not connect


Close and reopen EasySense, then switch the sensor off and on again. Check that it has not been paired through the device's Bluetooth settings.


Battery running low

Recharge using the supplied USB cable.


Still need help? Consult the full 1132 manual or contact Data Harvest technical support at 

support@data-harvest.co.uk.

For educational use. Read the full manual before use, particularly the safety and chemical-resistance guidance.